Block fields

Field type Field name EN Field name DE Patient data set Patient data item Procedure Step Warning Critical Comment Formula XML import
Link Import Importieren
Dropdown Device Gerät BIO - BIO - Show all device instances of Device type - BIO </XmlVersion><DeviceVersion>

<ProductName>

Time box Done at Erledigt am BIO MESD BIO - </Patient>

<AcquisitionTime>

Dropdown Lens status Linsenstatus BIO LST BIO - </Patient>

<LensState>

Dropdown LVC LSK BIO LVCST BIO - Lens State (treated, untreated- refsurgery)

TS: Laser Vision Correction/ Augenlaserkorektur [myopic,hyporopic]

<Patient>

<RefractiveSurgery>

Dropdown VBS GS VIT VBS BIO - Vitreous Body State <Patient><VitreousBodyState>
Number textbox Target refraction [D] Zielrefraktion [Dpt] TREF SPH BIO 0.25
Number textbox AL [mm] AL [mm] BIO AXL BIO 0.01 How to display status? </Patient>

<Al>

Number textbox ACD [mm] EXT VKT [mm] BIO ACD BIO 0.01 How to display status? </Patient>

<Acd>

Number textbox AQD [mm] INT VKT [mm] BIO AQD BIO 0.01 Make sure that units are4 converter, CCT should be in mm in calculation ACD-CCT
Number textbox LT [mm] LD [mm] BIO LNT BIO 0.01 Lens thickness

How to display status?

</Patient>

<Lt>

Number textbox CRW1 CRW1 BIO CRW1 BIO 0.01 100 * ((AL/K)'4 - (AL/TK)'4) + CCT score + K score
Number textbox CCT [µm] ZHD [µm] TOMO CCT BIO 1 >=400 and <480 <400 Central corneal thickness

How to display status?

</Patient>

<Cct>

Number textbox R1 [mm] R1 [mm] BIO --> KER K1(dim3) BIO 1 Maximal Radius <Keratometry><MaximalRadius>
Number textbox K1 [D] K1 [Dpt] BIO --> KER K1(dim1) BIO 1 >48 Check if calculation matches XML

Formula should be implemented, XML import data is for checking only

332/R1 (if KIndex is 1.332)

or 337.5/R1 (if KIndex is 1.3375)

<Keratometry>

<FlatK>

Number textbox Axis [°] Achse [°] BIO --> KER K1(dim2) BIO 1 <Keratometry>

<MaximalAxis>

Number textbox R [mm] R [mm] BIO --> KER KR(dim1) BIO 1 Radius ((R1+R2)/2)
Number textbox R2 [mm] R2 [mm] BIO --> KER K2(dim3) BIO 1 Minimal Radius <Keratometry>

<MinimalRadius>

Number textbox K2 [D] K2 [Dpt] BIO --> KER K2(dim1) BIO 1 Check if calculation matches XML

Formula should be implemented, XML import data is for checking only

332/R2 (if KIndex is 1.332)

or 337.5/R2 (if KIndex is 1.3375)

<Keratometry>

<SteepK>

Number textbox Axis [°] Achse [°] BIO --> KER K2(dim2) BIO 1 <Keratometry>

<MinimalAxis>

Number textbox K [D] K [Dpt] BIO --> KER KR(dim2) BIO 1 ((K1 + K2)/2)
Dropdown K Index K index BIO --> KER REFIND BIO - Refractive index <Keratometry>

<RefractiveIndex>

Number textbox ∆ K [D] ∆ K [Dpt] BIO --> KER CAST(dim1) BIO 1 Check if calculation matches XML

Formula should be implemented, XML import data is for checking only

K1-K2 @ A1 <Keratometry>

<DeltaD>

Number textbox Axis [°] Achse [°] BIO --> KER CAST(dim2) BIO 1 Axis from K1 <Keratometry>

<DeltaAxis>

Button Upload document Dokument hochladen BIO BIO Supports any file
Comment Comment Bemerkungen BIO COMM BIO -


Mockup